5.4
Micro-Ion Analog Output
Signal
A signal voltage proportional to the logarithm of the Micro-Ion pressure
indication is provided on the back of the electrometer module via a
standard 1/8 in. miniature phono jack.
Figure 5-4
Electrometer Analog Output Jack
5
9
1
6
Table 5-1
Remote Input/Output Pin Functions
Pin Number
Function
Pin Number
Function
1
Gauge On/Off Remote*
6
Degas On/Off Remote*
2
Ground
7
Ground
3
Not Used
8
Not Used
4
Gauge Status Common
9
Gauge Status N.O. (Normally Open)
5
Gauge Status N.C.
(Normally Closed)
*Active low inputs