Chapter 2
22
Series 358 Micro-Ion Controller Instruction Manual - 358013 - Rev. B
Convectron Gauge
Pressure Range
1 x 10
–4
Torr to 990 Torr, N
2
equivalent
Display Resolution
2 significant digits, except for 1 significant digit in 1 x 10
–4
Torr decade
Gas Type
N
2
, air (for direct reading)
Display Update Time
0.5 sec. typical. Switch selectable to 3 sec./reading, averaged
Analog Output
0 – 7 Vdc, logarithmic, 1 V/decade
Ion Gauge Turn-On Range
Less than or equal to 100 mTorr (1 x 10
-1
Torr)
Sensor Material
Gold-plated tungsten
Mounting Orientation
Gauge axis must be horizontal to provide accurate measurement above about
1 Torr
Operating Temperature
+4 °C to +50 °C ambient, noncondensing
Bakeout Temperature
+150 °C maximum, nonoperating, cable disconnected
Cable Bakeout Temperature
+105 °C maximum
Ion Gauge Pressure Range*
Table 2-3
Specifications