BG Series Goniometric Cradles
6.0 —
Connection to Non-Newport Electronics
6.1
Connections
WARNING
Newport is not responsible for malfunction or damage of BG
goniometric cradles when used with non-Newport controllers.
WARNING
Newport guarantees
“
”
compliance of BG goniometric cradles only if
used with Newport cables and controllers.
It is the customer's responsibility to modify the cable and take care of
sensor signal connections, when using the stage with non-Newport
controllers.
• The Mechanical Zero logic signal is open collector type. It supports up
to 30 V and 10 mA.
• The End-of-Run signal is open collector type with a 5.6 V protective
Zener diode.
Mechanical Zero:
I
in
max.: 10 mA
V max.: 30 V
NEWPORT
STAGE
End-of-Run:
I
in
max.: 16 mA
V max.: 5.6 V
NEWPORT
STAGE
If the “Mechanical Zero” output
is not used, a 1 k
Ω
/0.25 W
resistor must be connected
between pins #13 and #21.
5.6 V