Troubleshooting
In case where any caution/warning appeared
If any caution/warning message appeared on the touch
screen
and the display monitor, follow the
description below to take an appropriate action.
z
Caution/warning displayed on the touch
screen
and display monitor
Display
Measures
Touch
screen
Monitor
Valve Thermal
The valve's thermal switch for locking is
on. Wait 30 minutes before use.
No.01
Wait for about 30 minutes, and then use again.
Table Section Collision
The head plate is about to contact the
floor.
No.02
To continue the current operation, perform the following
operations.
•
Reverse Trendelenburg (
à
•
Back plate bending: Up (
à
•
Table up (
à
Table Section Collision
The seat plate is about to contact the
operating table.
No.03
To continue the current operation, perform the following
operations.
•
Lateral
tilt
in the return direction (
à
•
Trendelenburg in the return direction
•
Table up (
à
Table Section Collision
The leg plate is about to contact the
operating table.
No.04
To continue the current operation, perform the following
operations.
•
Leg plate bending: Up (
à
•
Trendelenburg (
à
•
Table up (
à
Motor Lock
The motor is locked and supplied with
overcurrent. Request repairs from
Mizuho or your distributor.
No.05
Request repairs from your distributor or Mizuho.
Power Supply
The AC power has broken down.
Request repairs from Mizuho or your
distributor.
No.06
Request repairs from your distributor or Mizuho.
Fuse
The fuse has blown out. It can
not
be
used by a battery mode. Request
repairs from Mizuho or your distributor.
No.07
Request repairs from your distributor or Mizuho.
Emergency Switch
The emergency stop switch was
pressed. Check the safety and then
deactivate the switch.
No.08
Check the safety and then release the emergency stop
switch.
Battery Discharged
The battery is out of charge. Charge the
battery.
No.09
Charge the battery.
Brake Lock
The surgical table cannot be locked.
Check if the emergency brake is
released.
No.10
Lock the emergency brake again.