background image

 

 
 

1/2 

 
 

MES-CK08-070-35EN-0 

PREPARED:  (2019-07-24) (Version 3) 

(2017-12-15) (Version 2)

 

Instruction for Use 

 

Trade Name: Carbon fiber extension 

 

 

Warning 

1.

 

When using operating table accessories, always pay 
close attention to the patient. Patients left in the 
same posture for an extended period of time are at 
risk for developing bedsores. 

2.

 

Avoid sudden impacts and excessive vibrations to 
the device. Doing so may cause unintended 
movement of the device and may compromise the 
device, may cause damage and/or harm. 

 

Contraindication/Prohibition 

1. This product is only validated for use with Mizuho 

operating tables. 

2. This product is only validated for use with operating 

tables shown in 

Specification

 1. Applicable models. 

3. Do not apply impermissible load on this product. 

(Doing so may damage this product.) 

Shape/Structure 

1. Shape/Structure 

 

122cm Carbon Fiber Extension 

 

Code No. 

Product Description 

08-070-35  122cm Carbon Fiber Extension   
08-070-36  50mm Pad for 122cm Carbon Fiber Extension   
08-070-37  70mm Pad for 122cm Carbon Fiber Extension   
08-070-38  90mm Pad for 122cm Carbon Fiber Extension   
08-070-39  110mm Pad for 122cm Carbon Fiber Extension   
08-083-67  70mm Mizuho Original Pad for 122cm Carbon 

Fiber Extension 

08-083-68  110mm Mizuho Original Pad for 122cm Carbon 

Fiber Extension 

 
2. Weight 

122cm Carbon Fiber Extension: 14kg 

3. Material 

 Carbon fiber extension: Carbon fiber reinforced plastic 

 Insertion holes: Stainless steel rod 

 Handle: Stainless steel rod 

 Mattress: Urethane foam 

 Support pole: Aluminum alloy rod 

 
 
 

Intended purpose 

This is an accessory to an operating table. This tool is used 
for supporting the back, head or legs of a patient. It is 
made of a carbon-fiber reinforced plastic material with a 
high degree of radiolucency. 

Specification 

1. Applicable models 

- MOT-5801S 
- MOT-5601KDB 
- MOT-5700 
- MOT-5701 
- MOT-3502 
- MST-7300B/BX 

2. Admissible load of the device 

122cm Carbon Fiber Extension: 135 kg   

Instructions for use 

1. When using for upper body, remove the head board and 

the back board from the operating table. 

2. When using for lower body, remove the leg section from 

the operating table. 

3. With the mounting pins of the operating table in the 

horizontal position, insert the carbon fiber extension 
completely over the pins. 

4. Pull the carbon fiber extension (main body) toward you 

to check that it is fully inserted and locked onto the 
operating table. 

5. Attach the specially-designed mattress.   

 

122cm Carbon Fiber Extension 

Caution 

Important Caution   
1. Be sure to check that the carbon fiber extension is fully 

inserted onto the extended pins of the operating table. 

2. When working with an operating table, take care not to 

allow this product to make contact or interfere with the 
table top or with other tools and appliances used in 
combination with this product. 

3.  The  attachment  brackets  of  the  device  are  made  of  a 

metallic material. Take the radiopaque brackets into 
consideration when using this product for X-ray 
photography. 

Storage/Life 

1. Do not store the device in high temperatures or in areas 

with high humidity where the temperature has drastic 
variations. 

Be sure to read this instruction manual before using this product. 
Keep this manual available for reference when needed. 

Unit: mm