background image

1/2 

 

MES-CK08-068-30EN-0 

PREPARED: (2022-10-18) (Version 3) 
 

Instruction for Use 

 

Trade Name: Back Section Board Attachment for Pediatric 

Warning 

1. When using operating table accessories, always pay close attention 

to the patient. Patients left in the same posture for an extended 
period of time are at risk for developing nerve paralysis and/or 
bedsores. 

2. Avoid sudden impacts or excessive vibrations to the device. 

Otherwise, the device may unintendedly rebound and vibrate that  
may harm the patient. 

 

Contraindication/Prohibition 

1. Do not exceed the allowable load specified in 

Specifications

, 3 

Allowable load. Otherwise, the product could break. 

2.  This product is only validated for use with Mizuho operating tables. 

Symbol mark for labeling 

  MD   

Medical Device

 

Specifications 

1.

 

Shape Unit: 

mm 

 

*Pad thickness: 08-068-32/08-083-63

 

70mm 

Code No. 

Product Description 

08-068-31 Back 

Sections Board Attachment for Pediatric *1 

08-068-31-R1 

Back Sections Board Attachment for Pediatric (R1) 

*2 

08-068-31-R2 

Back Sections Board Attachment for Pediatric (R2) 

*3 

08-068-32 

Pad for Back Sections Board Attachment for Pediatric 

08-068-33 

Radiolucent Top for Back Sections Board Attachment 

for Pediatric 

08-083-06 

Radiographic Table Tops Fixed Type for Back Section 

Board Attachment for Pediatric 

08-083-45 

Phenolic Board for Radiographic Table Tops Fixed 

Type for Back Section Board Attachment for Pediatric 

08-083-63 

Mizuho Original Pad for Back Section Board 

Attachment for Pediatric 

*1,*2 and *3: The size of mountable side rail is different. 

2.

 

Weight 

Main unit: 20 kg (Phenolic Board: 5.0kg)/Radiolucent Top: 5.0 kg/ 
Pad: 1.8kg 

3.

 

Allowable load 

50 kg 
 
 

4.

 

Material 

① Board: Thermosetting resin 
② 

Side rail: Stainless steel 

③ 

Lever: Carbon steel  

④ 

Frame:

 

Aluminum alloy  

⑤ 

Radiolucent Top: Thermosetting resin 

 Pad: Urethane foam 

⑦ 

Pad: Soft urethane foam

 

Note: The number corresponds to those used in 1. Shape. 

Intended purpose 

This is an accessory for an operating table. It is used for supporting a 
pediatric patient. 

Intended user 

This product is to be used by health care professionals, including but not 
limited to surgeons, nurses and biomedical technicians. 

Instructions for use 

 

Pad is attached to the Back Section Board Attachment for Pediatric. 

1. Attachment 

1-1. Remove the head plate and the back plate of the operating table. 
1-2. Check that the pins of the back plate are in a horizontal position, 

and insert the Back Section Board Attachment for Pediatric. 

1-3. Pull the Back Section Board Attachment for Pediatric toward you to 

check whether it is fully inserted. 

1-4. Attach the Radiolucent tops as needed. 
1-5. Attach the dedicated pad. 

 

2. Applicable operating tables 

  Refer to the brochure "OPERATING TABLE ACCESSORIES" of Mizuho. 

Warning/Caution 

1. Warning 

Ensure the attaching parts are secured. Loose condition can cause 
the product to come off or move, which may result in a patient injury. 

2. Important caution 

2-1. When working with an operating table, take care not to allow this 

product interfere with other accessories or the table top. 

2-2. For hygiene, be sure to use sterilized drapes on the areas on this 

product where the patient comes into contact with it. 

Storage/Life 

1. Do not store the device in high temperatures or in areas with high 

humidity where the temperature has drastic variations. 

2. Service life of this product: 7 years  

(Subject to the specified maintenance and inspection and is stored 
properly.) 

 
 
 
 
 
 
 
 
 

Be sure to read this instruction manual before using this product. 
Keep this manual available for reference when needed.  

Reviews: