5 External Input/Output Function
36
No. 99MBC109B
5.3
RS-232C Communication Function
By connecting to a PC or PLC via RS-232C, you can load measurement data
and manipulate various settings of the Counter through remote operation.
You can also save and load parameters.
5.3.1 Connections
Compatible plug and pin assignment
Compatible plug: D-sub 9-pin (female), inch screw type
Cable: A commercial RS-232C cross-type cable
2
RXD
IN
3
TXD
OUT
4
DTR
OUT
5
GND
-
6
DSR
IN
7
RTS
OUT
8
CTS
IN
1, 9
NC
-
1
2
3
4
5
6
7
8
9
1
2
3
4
5
6
7
8
9
Pin number
Pin number
Counter side
PC side
Pin number
Signal name I/O
1
5
6
9
Communication specification (conforming to EIA RS-232C)
Item
Description
Home position
DTE (Data Terminal Equipment), use a cross-type cable.
Communication
method
Half-duplex, non-procedural mode
Data transfer speed
4800 bps/9600 bps/19200 bps
Bit configuration
Start bit: 1
Data bits: 7 bits/8 bits (ASCII code, upper-case characters)
Parity: None/Even parity/Odd parity
Stop bit: 2
Tips
• Set communication conditions using parameters.
For details, see "3.3 Advanced Parameters" (page 14).
• Use commercial terminal software for communication with a PC.