DM Module Replacement
Most user and service adjustment data is stored on the
internal E2P Module. Before installing a replacement
DM Module. Remove the cover on both the old and
new modules.
Important:
Unplug the E2P Module from both DM
Modules. Insert the old E2P Module into the new
replacement DM Module. Refer to
Figure 6-10
for
the general location of the E2P Module
Figure 6-11
shows a close up of the E2P Module plugged in.
Figure 6-12
shows the E2P Module unplugged.
Installing the E2P Module in the new DM Module
insures only minimal adjustment is required after in-
stalling the replacement DM Module.
E2P MODULE
Figure 6-10: E2P Module Location