21
function module will be automatically shut off to save power.
Print stored data group
via Bluetooth.
In
the
memory
manager dialog box, select
“print”
will
the
selected
data
group.
Select “Print All” will print
all data groups stored
inside the instrument.
You can also print out
the test result directly after
each test.
On the main screen,
after each test, simple
press F1 to print out that
test result.
5.22 USB Communication
The instrument is equipped with a USB port on upper left of the instrument.
The PC can be connected with the instrument via the USB cable.
1. Lift the rubber flaps to uncover the connection port.
2. Insert the mini-USB end of the USB cable into the USB socket on the upside of main body.
3. Insert the other end into the USB port of the computer.
After installing the DataPro software and the USB driver, you can download the stored test data
from the tester.
The DataPro software helps manage and format stored data for high-speed transfer to the PC.
Data can be printed or easily copied and pasted into word processing files and spreadsheets for
further reporting needs. New features include live screen capture mode and database tracking.
Detailed information of the communication software and its usage refer to the software manual.
6. Measuring Technologies
6.1 Measuring Method
Single point measuring method: Put the probe to any point in the workpiece, the instrument
will show the probe located place thickness.
Two point measuring method: Using the probe to measure two times in the same point of the
tested piece, in two measuring, the probe parting plane keep 90º, the smaller value should be the
thickness of this point.
Multi-point measurement method: Taking times testing in an approximately 30 mm diameter
circular around, the minimum value is the thickness value of the tested piece.
Continuous measurement method: using single point measurement to take continuous
measurement along the specified path with less 5mm interval, the minimum value is the thickness
value of the tested piece.
6.2 Wall Measurement
During measuring, the probe parting plane can be along with the tube axis or vertical tube
axis. If meeting larger tube diameter, you should measure at the vertical axis. And when the tube