16
Press F1/F2 key to
move the highlight cursor;
Press
arrow
keys
to
increase/decrease
the
values.
Press
or F3 key to
confirm the setting.
Press
key to cancel the
change and exit.
5.8 Limit Setting
For the gauge, test
results beyond the limits will
be displayed with red color
to alarm the user. To
change the limit setting,
Press F1/F2 key to
move the highlight cursor;
Press
arrow
keys
to
increase/decrease
the
values.
Press
or F3 key to
confirm the setting.
Press
key to cancel the
change and exit
5.9 Changing Resolution
The
gauge
has
selectable
display
resolution, which is 0.1mm
and 0.01mm.
When the resolution is set to 0.01mm, the tested piece surface should be smooth to get
accurate test result. When measuring rough surface or coarse grained materials, it is suggested to
use low resolution.
5.10 Memory Management
5.10.1 Store a Record
By simply pressing the
key after a new measurement reading appears, the measured
thickness value will be saved to current selected data group. It is added as the last record of the
group.
5.10.2 View the Saved Record
This function provides the user with the ability to view the records in a desired data group
previously saved in memory. Following is the steps: