All the data from the walk test can be viewed in the following 6 screen shots and can
also be printed by following the instructions found in paragraph 4.2.
If the test results are printed, the paper report will only show the data related to the
“walking” stage of the test; please view the examples attached in this user manual.
Touch the STOP icon to end the test at any time.
WARNING
One of the parameters calculated during a walk test is the so-called Recovery
Time. The Recovery Time is defined as the time taken for the SpO2 value to
return to at least 99% of the average base value calculated during the initial
stage of the test.
3.13.2
Sleep Oximetry
To access the oximetry area from the main screen touch the
icon, next select the
icon to select the test.
This test records overnight patient parameter variations.
After approximately 5 minutes,
SPIRODOC
will enter standby (energy saving) mode; it
stops beeping and the display turns off. The led signal remains on. To check that the
device is functioning properly during standby mode, press
, after 1 minute
SPIRODOC
will return to standby mode.
Should the signal die during standby mode the device will automatically exit the
standby mode and a warning message will appear (sensor unplugged or finger not
detected correctly).
The data shown is the same as described in the previous test, except for information in
this present mode, which has not been envisaged for this test.
After the required time the test can be interrupted as previously described.
To print data see Paragraph 4.2.; an example of a test printout report can be found in
this user manual.