Obstructions or dirt on the sensor’s emitter and/or detector may cause a
sensor failure or inaccurate readings. Make sure there are no obstructions
and the sensor is clean.
Autoclaving, ethylene oxide sterilizing, may cause sensor damage. Do not
attempt to sterilize the sensor.
Unplug the sensor from
spirodoc
before cleaning or disinfecting to prevent
damaging sensor or device, and to prevent safety hazards for the user.
2.2.5
Device
WARNING
The maintenance operations detailed in this manual must be fully and
accurately carried out. If these instructions are not followed this may cause
measurement errors and/or an incorrect test interpretation.
Any modifications, adjustments, repairs or reconfigurations must be made
by the manufacturer or by personnel authorised by the manufacturer. Never
attempt to make a repair on your own. The set-up of configurable parameters
should only be made by qualified personnel. However, an incorrect set-up of
the parameters in no way endagers the patient
’s health.
High-
frequency emissions from “electronic” devices may interfere with the
correct operation of the device. For this reason, certain minimum clearances
(a few meters) should be observed when high-frequency appliances such as
a TV, radio, portable phone, etc. and other electronic units are operated at
the same time in the same room.
The device may give inaccurate readings if operated in the presence of
strong electromagnetic sources, such as electrosurgical equipment, or in the
presence of computed tomography (CT) equipment.
Do not use the device in the presence of magnetic resonance imaging (MRI)
equipment. MRI equipment may cause an induced current to the oximetry
sensor, resulting in patient injury.
If the PC and/or the printer connected to the
SPIRODOC
come into contact
with the area containing patient data, ref. directive EN 60601-1-1, it is
necessary that they conform to the directive EN 60601-1.
To dispose of the
SPIRODOC
, the accessories, any plastic consumable
materials (mouthpieces) as well as the battery, use only appropriate
containers or return all such parts to the dealer or to a recycling center. All
applicable local regulations must be followed.
If any of these rules are not followed then MIR will decline all responsibility
for any direct or indirect damages, however caused.
Use only the battery type indicated in the § Technical specifications.
Remove the battery from the device if the machine is not used for a long
period (several months).
The device may be powered through a PC by a USB cable. By this means, the
device works both on line with the PC, or individually powered by the PC.
Keep the device out of reach of children and of any person with mental
handicap.