Spirobank II

 

 

 

 

 

 

 

 

 

 

 

 

 

 

 

Rev.1.6 

Page 26 of 35 

User manual 

 
Once a test is selected the screen will display information concerning the type of turbine in use (reusable or disposable) including the 
necessary information to complete the test in the correct manner. 

 
To end a test press 

 key  

 

2.9.1

 

FVC test 

 

 

Proper execution of a FVC test must take into account the phases as described on the screen, more specifically: 
 

INSPIRE all the air 
EXPIRE fully with force 
INSPIRE fully with force 
 

It is possible (and may be helpful) to start the test by breathing at rest for a few moments. When ready to start

 inspire slowly as much air as 

possible

 (made easier by raising the arms wide apart) and then 

make a complete expiration as fast as

 

possible

. Then with the mouthpiece always 

held firmly in the mouth, complete the cycle by inspiring again as quickly as possible. This final inspiration may be left out if the inspiratory 
parameters (FIVC, FIV1, FIV1%, PIF) are not of interest. 

The optional initial inspiration phase can also be performed before inserting the mouthpiece in the mouth. 
After inspiring slowly and deeply, the following expiration must be made with the maximum effort by expiring all the air in the lungs as 
fast as possible. 
After 6 seconds of expiration the device will emit a continuous beep, this helps the user to understand whether the minimum expiry time 

has been reached, as recommended by the main international respiratory institutions. 
 

 

WARNING  

Accurate spirometry testing requires that the patient expire all the air in the lungs. 

 

The test may be carried out several times by repeating the cycle without taking the mouthpiece out of the mouth, in which case 

Spirobank 

II

 recognises the best test (largest FVC+FEV1) and will automatically display the results of the best test. 

To end the test press 

During the test the 

Spirobank II 

emits "beeps", the frequency of which are directly proportional to the inspired and expired velocity of 

the air. This helps the doctor understand when the velocity of the air is approaching zero, and the patient has almost exhausted all of the 
inspired or expired volume. 
In the maintenance section an explanation is given as to how this feature can also function as a very simple checking system  for the 

correct operation of the mobile “rotor” of the tu

rbine. 

For the FVC test to be judged as acceptable, besides breathing as deep as possible, it is also required that the forced expiratory time (FET) 
is sufficiently long to allow for the complete expiration of all air contained in the lungs. 
 

2.9.2

 

Test VC  

 

 

Ventilatory Profile 

The slow vital capacity test can be started by carrying out several breaths at tidal volume. After three or four such breaths an acoustic 
signal will be emitted to confirm that the ventilatory profile has been measured and that the patient may immediately proceed to perform 
the VC or IVC test. 
 

Expiratory Slow Vital Capacity: VC 

After the acoustic signal 

inspire slowly

 as much as air as possible and 

expire slowly 

as much air as possible. 

 

Inspiratory Slow Vital Capacity: IVC 

After the acoustic signal 

exspire slowly

 as much as air as possible and 

inspire slowly 

as much air as possible. 

To end the test press 

Follow the indications on the display carefully in order to carry out the test properly. 
 

2.9.3

 

MVV Test 

 

  

Not available in BASIC version 

Start  the  test  by  carrying  out  a  series  of  forced  inspirations  and  expirations  with  the  maximum  possible  amplitude.  The  suggested 
frequency is 30 breaths per minute. The test will end automatically after 12 seconds. 

 

WARNING 

To perform a test in  the service menu with the item “Set parameter”, at least one parameter related to this test must be 

activated/chosen, otherwise the icon will be disabled. 

Summary of Contents for Spirobank II

Page 1: ...Spirobank II Rev 1 6 Page 1 of 35 User manual Spirobank II User manual rev 1 6 Issue date 20 01 2021 Approval date 20 01 2021...

Page 2: ...HE Spirobank II 17 2 1 Switch on and switch off the device 17 2 2 Energy saving 18 2 3 Main screen 18 2 4 Symbols and Icons 18 2 5 Service menu 19 2 5 1 Reusable turbine calibration 22 2 6 Patient Dat...

Page 3: ...al packaging In the unlikely event that you have a problem with your device please use the original packaging and return it to the distributor or manufacturer Should this be the case please follow the...

Page 4: ...he patient by the doctor WARNING The manufacturer cannot be held responsible for any damage caused by the user of the device failing to follow instructions and warnings in this manual If the user of t...

Page 5: ...Spirobank II is continuously checked during manufacturing and therefore the product complies with the established security levels and quality standards laid down by the Council Directive 93 42 EEC for...

Page 6: ...spirometer are only to be used as a reference guide to purchase the correct size mouthpiece required These mouthpieces are clean but not sterile To purchase appropriate mouthpieces generally either pa...

Page 7: ...vent safety hazards for the user 1 2 5 USB connection cable Incorrect use or application of the USB cable may produce inaccurate measurements wich will show very inaccurate values of the patient s con...

Page 8: ...may be damaged or suffer a drop in overall performance The internal battery pack safety sensor could also be damaged as well by any of the above events Furthermore the user of the device could be har...

Page 9: ...device or the battery charger and do not use it any longer as any of these events may cause acid leakage overheating smoke breakage and or fire NOTE The battery pack includes an internal safety protec...

Page 10: ...ed against vertically falling drops of water Antenna symbol for devices that include RF transmitters FCC ID FCC Identification code indicating traceability to FCC compliance Rx ONLY Reference to US FD...

Page 11: ...ielectric breakdown ESD can be caused by a buildup of static electricity by tribocharging or by electrostatic induction At lower relative humidity as the environment is drier charge generation will in...

Page 12: ...nt to the next to ensure the maximum level of hygiene and safety For a correct interpretation of a spirometry test the measured values must be compared either to the so called normal or predicted valu...

Page 13: ...y Flow at 75 of FVC L s FEV05 Volume expired after 0 5 seconds L FEV05 FEV05 FVC x 100 FEV075 Volume expired after 0 75 seconds L FEV075 FEV075 FVC x 100 FEV2 Volume expired in the first 2 seconds of...

Page 14: ...usable soft sensor for adults Reusable paediatric soft sensor Range SpO2 Arms Range SpO2 Arms Range SpO2 Arms 70 100 1 19 70 100 1 470 70 100 1 390 70 80 0 554 70 80 1 626 70 80 1 851 80 90 1 32 80 90...

Page 15: ...octor 1 6 3 Oximetry alarms description Spirobank II is equipped with audio and visual alarm indicators to alert the operator to provide prompt patient attention or to abnormal device conditions Spiro...

Page 16: ...h of the above mentioned oximetry sensors is used 1 6 4 Other features Memory Memory capacity for over 10000 spirometric tests The precise number depends on the individual configuration so it cannot b...

Page 17: ...discharge protection 8kV contact 15kV air Magnetic field immunity 30 A m Radio Frequency Immunity 3V m 80 2700 MHz MIR will make available on request circuit diagrams component part lists description...

Page 18: ...displayed with a reduction of the indicators 2 3 Main screen On the main screen the following areas can be accessed patient data management area spirometry area oximetry area archive area archive are...

Page 19: ...e key and then the key The service menu shows the following list of items Change date time LCD settings Bluetooth settings Select language Delete memory Select standard Select predicted Select turbine...

Page 20: ...standard to be used ATS ERS or NHANES III with the keys and then press the setting takes effect and the device returns to the Service Menu WARNING If the NHANES III standard is selected it is not pos...

Page 21: ...hreshold or rise above the maximum threshold of the SpO2 and BPM values previously set Use keys and to decrease increase the values and select moving the arrow with the keys and At the end press to re...

Page 22: ...g of the turbine is necessary The disposable turbine is check before the application of the external packaging for this reason it doesn t requires a periodic calibration If a calibration must be perfo...

Page 23: ...robank II is calibrated correctly then the FVC syringe value will be 3 00 FVC x 1 026 BTPS 3 08 L FVC at BTPS If the ambient temperature is 20 C the FIVC syringe value will be 3 00 FIVC x 1 102 BTPS 3...

Page 24: ...en without modifying any data WARNING A new patient is not created from the previous patient when selecting this function Patient info however can be modified Future tests will be associated to the pa...

Page 25: ...efault settings of the PC software will be transferred to the device and will remain in the device even when used in stand alone mode for example when the Spirobank II is connected to the PC and the u...

Page 26: ...results of the best test To end the test press During the test the Spirobank II emits beeps the frequency of which are directly proportional to the inspired and expired velocity of the air This helps...

Page 27: ...rst screen displays a Flow Volume graph of the Forced Vital Capacity pressing it is be possible visualize the best value of FVC FEV1 FEV1 and PEF with the percentage change compared to the predicted v...

Page 28: ...argest FVC is 100 mL 2 11 Oximetry Testing WARNING Check if the oximetry function is available in the device this function is an option in some models This is not available in BASIC version WARNING Th...

Page 29: ...the Spirobank II timer resets itself and the device starts recording data For any type of oximetry test if the sensor is not properly connected the following message will be displayed on screen after...

Page 30: ...ct to biocompatibility tests Choose an application site on the patient s finger or toe where the light source will be directly over and in line with the detector The preferred sites are the forefinger...

Page 31: ...manage Spirobank II The user can perform on line wireless tests of spirometry or oximetry and the tests will be directly stored in the database of winspiroPRO software 3 2 PC connection via USB port W...

Page 32: ...liquid solution and shake it so as to remove any impurities Leave the turbine immersed for the time specified in the instruction of the solution WARNING To avoid irreparable damage to the reusable tur...

Page 33: ...an the sensor WARNING Do not sterilize by irradiation steam or by using ethylene oxide Unplug the sensor from the device before cleaning or disinfecting it The sensor included with the Spirobank II is...

Page 34: ...t was not performed correctly Repeat the test and follow closely the indications shown on the screen Certain spirometry and or oximetry parameters are not shown at the end of a test Personalized param...

Page 35: ...n If the product is utilised differently from the use described in the User s Manual If any alteration adjustment modification or repair has been carried out by personnel not authorised by the manufac...

Reviews: