Spirolab
Rev.1.8
Page 32 of 60
Adults
Children
USA
Knudson
ERS
Zapletal
MC-Barcelona
Zapletal
JRS-JAPAN
JRS-JAPAN
Chinese H.K.
Chinese H.K.
Pereira
Pereira
CECA1971
Knudson
NHANES III
NHANES III
GLI
GLI
Padilla-Perez
Knudson
If you choose the last predicted value NHANES III, the limitations provided for in the
past by the standard NHANES III will be applied.
WARNING
If you set the predicted value NHANES III are disabled the VC and MVV tests.
Delete archive
Using the numeric keyboard, enter the PIN set to permanently delete all the tests stored in
the device.
2.5.1
Calibrating turbines
WARNING
The turbine does not need calibration, just a periodical clean.
The disposable turbine is inspected before it is sealed in the package, and hence
does not need calibration.
If, however, if you want to carry out a calibration, bear the following in mind.
Calibration can be carried out on reusable and disposable turbines.
Calibration takes place based on the FVC (expiratory) and FIVC (inspiratory) values
measured during a test performed with a calibrated syringe.
To access the calibration setting, select the service menu item "Calibration";
the first
screen, "Last values", allows you to check current percentages applied by the device.