Table 12. Bits of status bytes description
Bit number of
<DATAE2>
status-bytes
Bit number of
<DATAE>
status-byte
Status word
compliance
11
Description
Recommendations
–
–
00
Normal mode, static temperature
–
0
0
10
Sensor is warming up.
Do not perform span calibration.
1
1
50
An abrupt signal change (due to
feeding a gas mixture to sensor
atmosphere) or increased noise
on sensor optical elements.
Do not perform span calibration or
zeroing. If this status longer than for
20 minutes at stable conditions,
replace sensor.
2
2
30
One of the signal values is lower
than it is allowed.
Possibility of condensed moisture on
optical elements. Dry sensor. After
drying, if this status lasts longer than
for 20 minutes at stable conditions,
replace sensor.
3
3
–
reserved
Does not contain any significant
information.
4
4
21
Dynamic
temperature
mode
(temperature
changes
faster
than by 0.6 °C/min).
Do not perform span calibration or
zeroing.
5
5
22
Dynamic
temperature
mode
(temperature
changes
faster
than by 2 °C/min).
Do not perform span calibration or
zeroing.
6
6
40
Exceeding temperature limits.
Check the ambient temperature.
7
7
90
Firmware failure (flash memory
issues).
Contact support for details.
8
–
11
Data request rate is more than 1
Hz.
Do not perform span calibration or
zeroing. Decrease request rate.
9
–
31
Zero shifts to negative value.
Possibility of condensed moisture on
optical elements. Dry sensor. After
drying, if this status lasts longer than
for 20 minutes at stable conditions,
perform zeroing (see Appendix D.1 for
procedure description).
10
–
1XX
Sensor operates in low power
consumption mode.
–
11
–
51
Complex status. Technological
failure.
Contact support for details.
12
–
–
reserved
Does not contain any significant
information.
13
–
–
reserved
Does not contain any significant
information.
14
–
–
reserved
Does not contain any significant
information.
15
–
–
reserved
Does not contain any significant
information.
(4
∨
5) & 9
–
24
Dynamic
temperature
mode;
zero shifts to negative value.
Do not perform span calibration or
zeroing.
11
Status word is contained in response to command <F> (see Table 17).