PR 5500 Instrument Manual
Device Installation
Sartorius EN-69
4.4.5 PR 5500/12 Digital Inputs and Outputs
The plug-in card has 4
passive opto-decoupled inputs
for process control.
The plug-in card also has 4 digital outputs with potential-free two-way contacts for process control.
The card is inserted in the Option 1/FB and/or Option 2 slot.
A max. of 2 PR 5500/12 plug-in cards can be installed!
Internal
connection
Contact strip
Digital inputs/
outputs
4 (CH1, CH2, CH3, CH4)
Input voltage
Logic 0: 0 to 5 V DC or open
Logic 1: 10 to 28 V DC
Passive, external power supply
required
Input current
<7 mA @ 24 V
<3 mA @ 12 V
Protected against incorrect polarity
Input frequency
max. 200 Hz (50% ratio)
Output
Two-way contact
Max. switching voltage
30 V DC/24 V AC
Max. switching current:
1 A
Switching
frequency
max. 0.5 Hz
Potential isolation
Inputs:
Yes, via optocoupler
Outputs:
Free relay two-way contact
External
connection
2x terminal, 6-pin
2x terminal, 4-pin
Wire gauge max. 1.5 mm
2
Cables
Screened
Connect the cable screen to the
device.
Cable length
max. 50 m
Dimens. (LxWxH)
60x106x22 mm
Weight
70 g
Summary of Contents for Sartorius Intec Maxxis 4 PR 5500