Extended Functions
PR 5500 Instrument Manual
EN-284 Sartorius
6 Extended Functions
6.1 System Maintenance
6.1.1 General Information
The following functions can be accessed via the
[System maintenance] menu.
When user management is activated, the logged in
user must have rights for the following:
-
Complete system maintenance
-
Importing (importing and restoring)
-
Exporting (exporting and saving)
6.1.2 Saving to selected Media (Backup)
This function is required to back up the current configuration and/or device database to connected storage
devices and/or network share connections.
This may be necessary for one of the following reasons:
-
The current data is needed on another device.
-
The configuration data is to be archived centrally.
-
The configuration and/or database might be deleted accidentally.
-
The configuration might be changed inadvertently.
The saved data can be restored later.
The backup stores the data sets in a format that can only be restored via the „Restore“ function.
The backup can be saved:
-
to the internal SD card,
-
to a USB stick, or
-
to the shared directories on the network.
Note
If ‚EXIT‘ key is pressed during backup stored data is deleted.