PR 5500 Instrument Manual
Commissioning
Sartorius EN-183
5.12.3 xBPI Parameter Input for Scale Function
The following parameters must be entered for this menu item:
-
Timeout for tare function depending on the application
-
SBN address for each xBPI scale, if bus operation is active
-
Serial number of xBPI scale or weighing module, if used in legal metrology
t
t
Press the [Paramet] soft key.
t
y
The following window appears.
t
t
Enter and confirm the following parameters.
[Settings locked]
The weighing point settings and assignment can
only be viewed as if the corresponding CAL switch
were closed; see Chapter 5.1.5.2.
[W&M]
Setting for operation in legal metrology. Select
[none], OIML, NTEP (for USA) or NSC (for
Australia); see Chapter 5.11.16.1.