In case of any mistake during the linearization process [CALE] (Calibration Error) will
be displayed.
7.3
[GEO] Geographical Factor Adjustment Procedure
The Geographical Adjustment Factor [GEO] is used to adjust the calibration based on the
current location. (Settings from 0 … 31 are available). Refer to table
to determine the
GEO factor that corresponds to your location.
Enter into Menu Mode:
Release the button.
The display shows the linearization weight value at 100% of scale capacity
[00150.0] (in this example, 150 kg).
13. Remove the weight after finish.
The scale is ready for weighing.
1.
Press and hold
Menu
M
+
until [M.E.N.U] displayed.
The 1
st
menu item [APPLIC] shown on display.
7 Calibration and Adjustment
Bench scale Puro®
Minebea Intec
EN-43