9 Maintenance/repairs/cleaning
9.1 Maintenance
The load cell PR 6211 is maintenance-free.
Corrosion protection grease must be applied to the load disc and the tip of the load cell.
Of-shore all-weather protection spray can be applied extensively to the load cell in
aggressive environments.
9.2 Repairs
The load cell PR 6211 is designed to be as robust as possible for the required measuring
accuracy and is highly reliable.
Should an electrical or mechanical defect nevertheless occur, the load cell must be
replaced.
Load cell repair is not possible.
9.3 Cleaning
Dirt on the load cell and movable parts of the scale must be cleaned as quickly as possible
-
if it inluences weighing, or
-
if it is corrosive to the cell or cable material.
NOTICE
Some cleaning agents may not be compatible with the load cell material.
When using cleaning agents, ensure that their compatibility with the load cell
material has been tested and approved (see Chapter
).
The free space around the load disc must be maintained.
9 Maintenance/repairs/cleaning
Compact Compression Load Cell PR 6211
Minebea Intec
EN-19