5.4 Equipotential bonding conductor
NOTICE
It is especially important that the ground is connected correctly to the components
and the cable junction box.
You must also ground the device separately and ensure the power supply is properly
shielded against the efects of lightning. Simply connecting the protective grounding
conductor is not enough!
If the installation is not carried out according to our instructions, this voids the
warranty. In particular, the entire installation, including the power supply, must be
suiciently protected against lightning.
In Ex areas, equipotential bonding must be established to avoid any compensatory
currents between the individual conducting system components.