MP 26/10
Installation and Calibration Manual
Minebea Intec
EN-32
5.3.1.
Analog output configuration example
For example, the device is calibrated for 50% of its measurements in 5000 g, O.typ = 2 (4 to
20 mA), Out.0 = 500, Out.1 = 2500, O.src = 7, O.FAI = 0
Now, the analog output generation for the weight on the system is
-357 g
0 mA
350 g
2.8 mA
500 g
4.0 mA
1500 g
12.015 mA
2500 g
20 mA
3000 g
24.039 mA
EEEE
25.128 mA
Note :
Out.0 and Out.1 have to be entered as weight values.
In case of a weight greater than the Out.1, the device produces 12 V or 25 mA
In case of a weight lower than the Out.0, the device produces 0V or 0 mA/<4 mA
If there is an input value error, then the output behavior for this can be determined using O.FAI.
The smaller the span, the higher the effect of input variations and resolution.
Using current and voltage output in parallel is possible only in galvanically isolated circuits.