MP 26/10
Installation and Calibration Manual
Minebea Intec
EN-31
5.3.
Output Out.3 (analog)
The device can be configured to take the output measure as constant voltage or current.
The procedure for configuring Out.3 is:
1.
Go to configuration
Out.3 and set/unset the required values using the down arrow
button by clicking the value field
Name
Value
Description
O.tYP
Signal type OUT3
1
0 ... 20 mA continuous
2
4 ... 20 mA continuous
3
0...10 V continuous
4
2...10 V continuous
Out.0
-1999...9999 Scaling of entered value to 0% of analog output type
(Out.0 =entered_value0 =0% )
Out.1
-1999...9999 Scaling of entered value to 100% of analog output type
(Out.1 =entered_value1 = 100% )
O.Src
Signal source for analog output OUT3
3
Process value (scaled and corrected value ie displayed value)
7
Measured value (raw value of bridge signal)
O.FAI
Fail behaviour
0
upscale
1
downscale