![Minebea Intec Maxxis 5 Original Installation Manual Download Page 97](http://html1.mh-extra.com/html/minebea-intec/maxxis-5/maxxis-5_original-installation-manual_1801172097.webp)
Example:
Connection for PR 5900/13: Voltage output
①
2.2 k
Ω
/1 k
Ω
with 24 V/12 V
②
Screen clamping rail or cable gland
③
U
DC
= 24 V 0.5 A power supply unit
When the output is active (true), the output voltage drops from 24 V/12 V to 0 V. The
load resistance must be 2.2 k
Ω
/1 k
Ω
.
4.6.7
PR 5900/17 digital inputs and outputs
The plug-in card has 6 passive opto-decoupled inputs and 8 passive opto-decoupled
outputs for process control.
All inputs and outputs have a common GND (-).
The card is inserted in the option 1 and/or option 2 slot.
A maximum of 2 PR 5900/17 plug-in cards can be installed.
4 Device installation
Process Controller Maxxis 5 PR 5900
Minebea Intec
EN-95
Summary of Contents for Maxxis 5
Page 137: ......