[START + LEVEL↑]
By tapping [START], the detection of level is started.
Graph starts drawing when the weight value exceeds the set
value of [GRAPH TRIGGER LEVEL], and draws once until the
time of [T].
[START + LEVEL↑↓]
By tapping [START], the level detection starts.
Graph starts drawing when the weight value has passed the
set value of [GRAPH TRIGGER LEVEL], and draws once until
the time of [T].
[LEVEL↑]
By tapping [START] or when the weight value exceeds the set
value of [GRAPH TRIGGER LEVEL], graph starts drawing and
draws once until the time of [T].
[LEVEL↑↓]
By tapping [START] or when the weight value has passed the
set value of [GRAPH TRIGGER LEVEL], graph starts drawing
and draws once until the time of [T].
[BATCH START LINKED] By tapping [START] or by inputting [BATCH START], graph
starts drawing and draws once until the time of [T].
[SINGLE] is set as default.
9.18 Graph trigger level setting (GRAPH TRIGGER LEVEL)
The level of graph trigger is set.
The setting is inputting with ten keys in the range from -999 999 to 999 999 in every 1.
[0] is set as default.
9.19 Hold of comparative result (COMPARATIVE RESULT HOLD)
Hold operation of the comparative result that synchronizes with hold signal (refer to
Chapter
) is selectable from [INVALID] or [EFFECTIVE].
[INVALID] is set as default.
The targets of the hold are the signals of [FULL FLOW] (Full low), [MEDIUM FLOW] Middle
low, [DRIBBLE FLOW] (Fine low), [OVER], [OK], [UNDER], [FULL] and [ZERO BAND]
(Empty).
Note:
During the hold of comparative result, the status data of status display, signal output,
standard RS485 interface, RS232C interface and RS422/485 interface corresponding to
the hold target of comparative result is held also.
9.20 Hold of 2-pin method serial interface (S-I/F HOLD)
Hold operation of the 2-pin method serial interface (S-I/F) outputs data that synchronizes
with hold signal (refer to Chapter
) is selectable from [INVALID] or [EFFECTIVE].
[INVALID] is selected as default.
The holding target are [GROSS], [NET], [TARE VALUE], [GROSS STATUS], [NET STATUS],
[TARE STATUS], [STABLE], [CENTER ZERO], and [PRESET TARE].