— CHECK 1/1
Check screen 1/1
— ROM VERSION
see Chapter
— CHECK OF DISPLAY
see Chapter
— CHECK OF I/O
see Chapter
— MONITOR
see Chapter
— BCD OUTPUT
see Chapter
— ANALOG OUTPUT
see Chapter
— MAINTENANCE DATE
see Chapter
— MAINTENANCE NO
see Chapter
CHECK
20.2 Customer service
Warranty
The device warranty period is one year from the date of delivery.
Consult Minebea Intec sales oice or point of purchase for repairs or customer service
during the warranty period.
Repair
Before requesting a repair, check once again that all of the wiring/settings/adjustments
are correct.
In particular, check the load cell for disconnected and/or broken wires. If the problem
persists, contact Minebea Intec sales oice or point of purchase to request repair service.
20.3 Speciications
20.3.1
Analog speciications
Bridge power supply
Within DC12 V±0.3 V, 140 mA (DC3 V and 6 V can be switched) with
remote sensing applied.
Applicable load cell
Up to 4 load cells (for 350
Ω
) can be connected in parallel.
Input sensitivity
0.3 µV/d or more (d = minimum scale interval)
Input range
-3.1 mV/V to 3.1 mV/V
Zero setting range
±2.5 mV/V
Non-linearity
0.01 % F.S.
Zero
±0.1 µV/°C (calibrated at an input sensitivity of 0.3 µV/d or more)
Temperature inluence
Sensitivity
±0.000 8 % F.S./°C (calibrated at an input sensitivity of 0.3 µV/d
or more)
Input noise
±0.2 µVp-p or less (at the default setting for analog ilter, digital
ilter, stabilization ilter and A/D sampling rate)
Analog ilter
Approx. 4 Hz (can be changed to 2 Hz, 6 Hz, 8 Hz, 10 Hz or 100 Hz)
A/D sampling rate
200 times/s (can be switched to 1 000 times/s, 100 times/s, 50
times/ s, 20 times/s or 10 times/s)
A/D internal resolution
24 bit
20 Appendix
Weight Controller CSD-918
Minebea Intec
EN-209
Summary of Contents for CSD-918 Series
Page 239: ......