Example 2
Net value cancelled 300 g tare weight set in Example 1
Scale interval 0.1 g
Scale interval 0.5 g
Scale interval 1 g
0 g
Tare weight
300 g
600 g
1000 g
–300 g
0 g
600 g
600 g
700 g
Net Value
weighing value
Gross Value
The weighing values from –300 to 300 g becomes the irst range (scale interval 0.1 g).
The weighing values from 300 to 600 g becomes the second range (scale interval 0.5 g).
The weighing values from 600 to 700 g becomes the second range (scale interval 1 g).
Note:
-
When only used only within the irst range (C function CF-40: OFF), it is not
necessary to set the range.
-
Ensure that the scale interval of each range conforms to irst range < second range <
third range.
-
Ensure the boundary value of each range conforms to second range ≤ third range ≤
maximum load.
-
The range upper boundary within the second range is the maximum load to switch
to two ranges.
The range upper boundary within the third range is the maximum load to switch to
three ranges.
-
The comparator setting and the preset tare weight cancellation setting become the
scale interval for the irst range.
8.15 Stability detection duration during calibration
Stability detection duration calibration is set by C function CF-97.
Setting range: 00 to 99 (Unit: 0.1 s
The default is [05].
8.16 Digital linearization clear
Digital linearization compensation data can be cleared by C function CF-99.
[CF-99]
[LNR.CLR]
1. If the [F] key is pressed with[CF-99] displayed, [LNR.CLR] blinks.
2. Press the [ON-OFF] key to discontinue digital linearization.
The display returns to stand-by condition, and digital linearization is not cleared.
Weight indicator CSD-903
8 Various operations by C function data
EN-64
Minebea Intec