The A/D load cell input is negative overload.
All lights
Blinks if accumulation value or accumulation times exceed the
display range.
Battery LO error. Contact Minebea Intec.
Backup data corrupted. Contact Minebea Intec.
20.2.2 Error sub-display
Error code
Contents
Actions
ZERO set is executed out of
range.
Execute ZERO set within its
operational range.
TARE WEIGHT is executed out
of range.
Execute TARE WEIGHT within
its operational range.
The accumulation value and
the accumulation number
exceed the display range.
Clear the accumulation count
when the accumulation display
blinks.
Note:
Sequence error sub-display blinks until sequence error is released.
Press the [F] key to release sequence error.
20.2.3 Sequence error display
Error code
Error
No.
Contents
Actions
1
A forced temporary stop-
page has occurred due to
inputting a temporary stop
while weighing.
Check the cause of the
temporary stoppage and
restart
3
The load value is insui-
cient, even if post-
batching is executed.
Change the time and ma-
ximum times of post-
batching and restart
4
The comparison values are
contradictory.
Check the settings and
resolve the contradiction
5
Inlow time exceeds the
time limit.
Check the inlow gate for
obstructions
6
Discharge time exceeds
the time limit.
Check the discharge gate
for obstructions.
20 Troubleshooting
Weight indicator CSD-903
Minebea Intec
EN-231