— CAL
Initiates a span calibration procedure (zero and
span)
— LIN
Initiates a linearity calibration procedure (zero,
mid-point and span)
— GEO
Geographical Adjustment Factor (GEO) is used
to adjust the calibration based on the current
location. (Settings from 0 … 31, default 12)
— END
Leave menu level
7
Calibration and Adjustment
Enter this menu to perform calibration and adjustment of your scale.
Initial Calibration
When the scale is operated for the first time, a Span Calibration is recommend to ensure
accurate weighing results. Before performing the calibration, be sure to have the
appropriate calibration weights.
Adjust the GEO setting according to your location (see table
).
CAL.ADJ
7.1
[CAL] Calibration Procedure
Enter into Menu Mode:
Release the button.
1.
Press and hold
Menu
M
+
until [M.E.N.U] displayed.
The 1
st
menu item [APPLIC] is displayed.
2.
Press the
Unit
button move to next menu item or
Mode
F
button move to previous
menu item.
3.
Change menu item until [CAL.ADJ] is displayed.
4. Press
Off
(Yes) button to enter into sub-menu item.
Sub-menu item [CAL] (Calibration) is displayed.
7 Calibration and Adjustment
Compact Scale Puro® - Basic
Minebea Intec
EN-27
Summary of Contents for Compact Scale Puro Advanced EF-P1
Page 46: ...all dimensions in mm inch Compact Scale Puro Basic 12 Technical data EN 44 Minebea Intec...
Page 48: ...LargeFlat LF Compact Scale Puro Basic 12 Technical data EN 46 Minebea Intec...
Page 51: ...all dimensions in mm inch 12 Technical data Compact Scale Puro Basic Minebea Intec EN 49...
Page 55: ...all dimensions in mm inch 12 Technical data Compact Scale Puro Basic Minebea Intec EN 53...