— S-SQMIN
Activates display or GMP-compliant printout
SM)
— SQMIN 1
SQmin 1
— SQMIN 2
SQmin 2
— DISPLAY
SQmin value display
— NO
No*
— YES
Yes
— GMP PRT
GMP-compliant printout
— NO
No*
— YES
Yes
substances are weighed with the highest degree of accuracy for volume determination.
This additional function ensures that weighing results lie within deined tolerance limits
corresponding to the requirements of your quality assurance system.
System Requirements
The scale must be set up by a service technician to be able to use the SQmin function. The
technician will determine the permitted minimum sample quantity and load this to your
scale using the guidelines of your QA system. He will document this setting via a
"Weighing module test as per USP" certiicate in which the measurements and min.
sample quantity are logged. The SQmin function ensures that the weighing results
correspond to USP guidelines. These SQmin settings cannot be changed by the user.
Features
-
Displaying the minimum sample quality: The value is displayed in the text line for
4 seconds after pressing the
key.
-
If the minimum sample quantity has not been reached: The
symbol is displayed.
Weight values are marked with a "!" in the printout.
-
GLP header: The minimum sample quantity entered for "SQmin" can be included on
the printout.
Setting SQmin parameters
The SQmin display must be turned on to use the SQmin function.
Menu: [SETUP] - [SQMIN] (see chapter
SQMIN
* Factory setting
7.4.1
Example: SQmin Operation
Example:
Determining sample weights while monitoring the minimum sample quantity
(in this example, SQmin = 100 g).
Coniguration: The SQmin display must be turned on.
1. Place the container for the sample on the scale and tare with
key.
2. Place th esample on the scale.
The minimum sample quantity is not reached (symbol
⚠
).
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
7 Operating
EN-94