The key
must be assigned the function "Set preload" (menu item [SET.VORL.]
1.9.8) (see Chapter
menu item [CAL.JUST] 1.9).
Note:
After linearization, the
key must be reassigned its original function, e.g. External
calibration/adjustment with default weights (see Chapter
menu item [CAL.JUST]
1.9).
Procedure
5.6.8
Delete preload
Setup information
-
The key
must be assigned the function "Delete preload" (menu item
[CLR.VORL.] 1.9.8) (see Chapter
menu item [CAL.JUST] 1.9).
Note:
After linearization, the
key must be reassigned its original function, e.g. External
calibration/adjustment with default weights (see Chapter
menu item [CAL.JUST]
1.9).
1. Zero the weighing platform by pressing
.
2. Place the preload weight onto the weighing platform.
3. Start "Set Preload" with key
.
After a short time, the preload is adopted and the automatically switches to
weighing mode.
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
5 Getting started
EN-56
Minebea Intec