— CAL.JUST
1.9 - Calibration, adjustment
— CAL.EXT.
1.9.1 - External calibration with default weight
— CAL.E.AUT.
1.9.2 - External calibration, weight is detected
(see Chapter
menu item 1.18.1)
[HND.EXT.G]- [CAL.JUST]
— CAL.E.USR.
1.9.3 - External calibration with user weight*
— CAL.INT.
1.9.4 - Internal calibration (IS scales only)
— LIN.INT.
1.9.5 - Internal linearization (IS scales only)
1)
— LIN.EXT.
1.9.6 - External linearization with standard
weights
1)
— LIN.E.USR
1.9.7 - External linearization with user weights
1)
— SET.VORL
1.9.8 - Set preload
— CLR.VORL
1.9.9 - Delete preload
— GESPRT.
1.9.10 - Key locked
Menu tree for assigning functions to the key
Menu: [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.JUST] (1.9) (see also
Chapter
* Factory setting
1)
Not possible for veriied scales
5.6.6
External linearization
Setup information
-
External linearization on scales used in legal metrology is only possible when the
locking switch is open (see Chapter
).
-
The
key must be assigned the function "external linearization" (menu item
[LIN.EXT.] 1.9.6 or [LIN.E.USR] 1.9.7) (see Chapter
menu item [CAL.JUST] 1.9).
Note:
After linearization, the
key must be reassigned its original function, e.g. External
calibration/adjustment with default weights (see Chapter
menu item [CAL.JUST]
1.9).
Procedure
Menu: [SETUP]- [WP-1]- [INTERN.]- [PARAM.1]- [CAL.JUST] (1.9) (see also
Chapter
1. For scales used in legal metrology: Open the locking switch (see Chapter
2. Zero the weighing platform by pressing
.
3. Set service mode (see Chapter
).
4. Start linearization with the
key.