5 Getting started
5.1 General information
You can use the Combics 1 to record weight values and display weight values.
First, use the menu to conigure the indicator for the desired application (printer settings,
etc.). Then you can begin weighing.
The indicator keypad is used for operation (see chapter
a weighing mode function and another function in the menu. Some of the keys also have
an additional function when pressed and held for longer than 2 seconds.
When a key is pressed that does not have an active operating mode function, an
acoustical signal
1)
(double beep) sounds and the message "-------" is displayed for
2 seconds. The display then returns to the previous screen content.
1)
only if menu item [SIGNAL] 8.2 is switched on (see chapter
).
5.2 Switching on the device
Whenever it is switched on, the device performs a self-test. This will display all display
segments for a few seconds.
The scale starts in measurement mode. You have to open the menu mode in order to
apply settings (see Chapter
).
5.3 Warm-up time
To provide accurate results, the device requires a warm-up time of at least 30 minutes
after connection to the mains. Only then will the device have reached the necessary
operating temperature.
Using a veriied device in legal metrology
Allow a warm-up time of at least 24 hours after connection to the mains.
To switch on the scale, press the key
.
The scale is started in the status in which it was last switched of.
Then the display for the measurement mode appears.
Combics 1 Series Complete Scales Models CAW1P, CAW1S, CAS1
5 Getting started
EN-44
Minebea Intec