Product Shipment
Shipping containers should be inspected for damages, and the "shock watch" and "tilt
watch" indicators for signs of mishandling during shipment. If any problems are found,
make notes of any discrepancies and immediately report to the shipping carrier and to
Mindray/Zonare
’s shipping department representative.
All installation and set-up of equipment should be done following this official
Mindray/Zonare product installation procedure.
Electrical Requirements
Medical Grade Receptacles grounded (3-Prong) 110-120 VAC, 15 amp, 60Hz. wall
outlets (or proper 220-240 VAC outlets for systems configured for these power
requirements) are required for the ZS3 ultrasound system.
Environmental and Space Requirements
The ZS3 ultrasound system may be located anywhere within the facility, as long as
there is appropriate power outlets and remote network connectivity (at facilities utilizing
a DICOM print/storage device hook-up).
Uncrating
1. Prior to opening any packaging, inspect the "shock watch" and "tilt watch"
indicators (attached to the outside of the main shipping box) for signs that the
system has been subjected to shock or tilt conditions during shipment.
2. Also inspect the shipping container for any visual signs of rough handling or abuse
during shipment.
3. If any discrepancies from shipment are noted, follow the instructions listed in the
“Product Shipment” section above.
4. Remove all banding straps surrounding the cardboard main shipping container.
Summary of Contents for Zonare ZS3
Page 1: ...ZS3 Diagnostic Ultrasound System Service Manual ...
Page 120: ...ZS3 Service Manual Page 120 of 295 FTP Setup Enters Setup Network FTP ...
Page 124: ...ZS3 Service Manual Page 124 of 295 ...
Page 131: ...ZS3 Service Manual Page 131 of 295 13 System Diagrams ...
Page 132: ...ZS3 Service Manual Page 132 of 295 Power Block Diagram Figure 13 1 ZS3 Power Block Diagram ...
Page 133: ...ZS3 Service Manual Page 133 of 295 Cabling Diagram Figure 13 2 ZS3 Cabling Diagram ...
Page 138: ...ZS3 Service Manual Page 138 of 295 Figure 14 6 ZS3 ...
Page 185: ...ZS3 Service Manual Page 185 of 295 17 Preventative Maintenance Forms ...
Page 217: ...ZS3 Service Manual Page 217 of 295 Figure 18 8 ZS3 ...
Page 252: ...ZS3 Service Manual Page 252 of 295 Figure 19 43 ZS3 Power Cable USB Cable ...
Page 295: ...P N 046 014026 00 2 0 ...