ZS3 Instructions for Use
45
|
P a g e
TEE/Laparoscopic Transducers
The ZS3 system will support laparoscopic imaging when used with the C12-4lp transducer, and
TEE imaging when used with the P8-3 TEE and P8-3m TEE transducers.
Refer to TEE Transducer and laparoscopic transducers for instructions on
inspecting, cleaning, and maintaining the TEE and laparoscopic transducers.
C12-4lp
Figure 4-1: Laparoscopic Transducer Handle and Adjusting Levers
Scan plane rotation is controlled by two adjusting levers on the control handle and can be
rotated from -90° to 90°. The adjusting levers control the deflection of the tip. The left adjusting
lever is used to control the transducer Up/Down (both 90°). The right adjusting lever is used to
control the transducer Left/Right (both 90°).
P8-3 TEE
Scan plane rotation is controlled by two push-buttons on the control handle and can be
rotated from 0° to 180°. The wheels on the handle of the transducer control the deflection of
the tip. The smaller lower wheel is used to control the transducer Left/Right tip deflection
(
≥45° left and ≥45° right). The larger wheel on top of the handle is used to control the
transducer Anterior/Posterior tip deflection (
≥120° upwards and ≥90° backwards). Both
wheels have a friction lock and freely moving mode.
Left adjusting lever
Right adjusting lever
Summary of Contents for Zonare ZS3
Page 1: ...I ZS3 Ultrasound System Instructions for Use ...
Page 6: ......
Page 12: ......
Page 38: ......
Page 74: ......
Page 78: ......
Page 86: ......
Page 110: ......
Page 128: ......
Page 146: ......
Page 150: ......
Page 158: ......
Page 160: ......
Page 164: ...ZS3 Instructions for Use 147 P a g e ...
Page 165: ...P N 046 012347 01 3 0 ...