ZS3 Instructions for Use
144
|
P a g e
22.
Specifications
Type
Parameter
Value
Electrical
Power requirements
100–240V~, 50–60Hz, 6A max
Power consumption (no peripherals) (max.)
180W (616 BTU/hr.)
Power consumption (w/peripherals) (max.)
470W (1608 BTU/hr.)
Environmental
(operating)
Cooling requirements
See power consumption above.
Air temperature
0–35°C (32–95°F)
Humidity
15–80%, noncondensing
Pressure
700–1060 hPa
Environmental
(storage and
transportation)
Air temperature
-20–60°C (-4–140°F)
Humidity
15–90%, noncondensing
Pressure
500–1060 hPa
Physical
(weight)
ZS3 system
≤
79.5Kg (not including peripherals)
Physical
(dimensions)
Height, min (display lowered for transport)
106 cm (42 in.)
Width
51 cm (20.1 in.)
Depth
72 cm (28.2 in.)
I/O connectors Ethernet (1 port)
RJ-45 – 10/100BaseT
USB 2.0 (3 ports)
USB-Type A (VBUS 5.0V, 0.5A max)
External Video (1 port)
HDMI 1280x1024
Serial port
eSATAp
Wireless Option
Manufacturer
B&B Electronics / Quatech
Summary of Contents for Zonare ZS3