ZS3 Instructions for Use
89
|
P a g e
Map Position
Move the color bars onscreen Up and Down the screen to prevent interference with
imaging data.
RLB Index
On/Off
Turn display of RLB Index On and Off screen.
M-STB Sensitivity
0-4
Allows user to adjust the degree of sensitivity used when displaying M-STB Index.
Higher values are more sensitive to motion and will result in more frequent changes to
onscreen graphic.
Select Pane
Right or Left
Toggle to select an imaging Pane before making image pane specific changes (like
selecting data to display etc)
Map
Toggle to alternate 2D SWI color maps.
Map 0 indicates no color overlay. 2D SWI data is displayed in grayscale.
Map 2 and 3 are color maps
Scale
Change the maximum m/s or kPa displayed on the 2D SWI color bar.
Invert
Toggle to invert the 2D SWI color map.
Tissue Tracking Quantitative Analysis (TTQA)
CAUTION:
It is provided for reference, not for confirming a diagnosis.
Tissue Tracking Quantification Analysis (TTQA) is a separately purchased option
which
provides
a tissue tracking
function for myocardial
movement evaluation.
The ultrasound system scans each pixel position frame by frame within the cardiac cycle, and then
uses a regional matching method and auto-correlation searching method to trace each spot and
calculate the movement, to determine myocardial motion in a more quantitative method.
TTQA
must be assigned to a programmable function key before use. These keys can be configured
in
Setup/Keys
.
4. Press
Setup / Keys
.
5. Click to select
TTQA
to the desired key.
6. Click Save to complete the setting.
Once configured,
TTQA
is invoked by pressing the configured function key.
•
Basic Procedures for TTQA
1.
Select and open a Cardiac Clip from Review or Archive, with ECG containing at least 2
cardiac cycles (preferably 6 cardiac cycles)
2.
Press the configured key (such as F2) to enter the
TTQA
feature.
a. System enters the TTQA function with 1 heartbeat playing on the upper screen and
TTQA strip on lower screen
3.
Select which view is to be evaluated using the View Type OLED
4.
Select desired Tracking Cycles (1 or 3) using the OLED
5.
Select which heart beat range will be evaluated using the Cycle OLED
Summary of Contents for Zonare ZS3
Page 1: ...I ZS3 Ultrasound System Instructions for Use ...
Page 6: ......
Page 12: ......
Page 38: ......
Page 74: ......
Page 78: ......
Page 86: ......
Page 110: ......
Page 128: ......
Page 146: ......
Page 150: ......
Page 158: ......
Page 160: ......
Page 164: ...ZS3 Instructions for Use 147 P a g e ...
Page 165: ...P N 046 012347 01 3 0 ...