8 - 11
3.
Set
Averaging
.
8.6.8
Changing the Sensitivity (for Mindray SpO
2
)
The SpO
2
value displayed on the monitor screen is the average of data collected within a
specific time. The shorter the averaging time is, the quicker the monitor responds to
changes in the patient’s oxygen saturation level. Contrarily, the longer the averaging
time is, the slower the monitor responds to changes in the patient’s oxygen saturation
level, but the SpO
2
measurement is more stable. For critically ill patients, selecting
shorter averaging time will help understanding the patient’s state.
To set the averaging time, follow this procedure:
1.
In
Continuous Monitoring
mode, select the SpO
2
numeric area or waveform area
to enter the
SpO2
menu. In
Spot Check
mode, select the
Main Menu
quick key
→
from the
Parameters
column select
Setup
→
select
SpO2
.
2.
Select the
Setup
tab.
3.
Set the
Sensitivity
.
8.6.9
Showing/Hiding PI
You can set whether to display PI in the SpO
2
parameter area. To do so, follow this
procedure:
1.
In
Continuous Monitoring
mode, select the SpO
2
numeric area or waveform area
to enter the
SpO2
menu. In
Spot Check
mode, select the
Main Menu
quick key
→
from the
Parameters
column select
Setup
→
select
SpO2
.
2.
Select the
Setup
tab.
3.
Switch on or off
Display PI.
NOTE
•
This function is only available for Mindray SpO
2
and Masimo SpO
2
.
8.6.10
Monitoring SpO
2
and NIBP Simultaneously
When monitoring SpO
2
and NIBP on the same limb simultaneously, you can switch on
NIBP Simul
to lock the SpO
2
alarm status until the NIBP measurement ends. If you
switch off
NIBP Simul
, low perfusion caused by NIBP measurement may lead to
inaccurate SpO
2
readings and therefore cause false physiological alarms.
In
Continuous Monitoring
mode, you can set the
NIBP Simul
. Follow this procedure:
1.
Select the SpO
2
numeric area or waveform area to enter the
SpO2
menu.
2.
Select the
Alarm
tab.
3.
Set
NIBP Simul
.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...