A - 16
A.6.5
CO
2
Specifications
Standard
Meet standard of ISO 80601-2-55
Technique
Infrared absorption
Apnea delay
10 s, 15 s, 20 s, 25 s, 30 s, 35 s, 40 s
Measurement
range
0 to 150 mmHg
Accuracy
Full accuracy mode:
0 to 40 mmHg: ±2mmHg
41 to 76 mmHg: ±5% of reading
77 to 99 mmHg: ±10% of reading
100 to 150mmHg: ±(3mmHg + 8% of reading)
>150mmHg CO
2
: unspecified
ISO accuracy mode:
Add ±2 mmHg to the full accuracy mode
Accuracy drift
Meet the requirement for measurement accuracy within 6 hours.
Resolution
1 mmHg
Sample flowrate
Connecting the Oridion sampling line: 50 ml/min
Sample flowrate
tolerance
±15% or ±15 ml/min, whichever is greater.
Start-up time
20 s (typical), 90 s (maximum)
Response time
Measured with an Oridion sampling line: ≤5 s @ 50 ml/min (standard
sampling line) or ≤6.5 s @ 50 ml/min (extended sampling line)
Measured with a DRYLINE
TM
PRIME sampling line: ≤5 s @ 50 ml/min
Rise time
Measured with an Oridion sampling line: ≤250 ms @ 50 ml/min
(standard sampling line) or ≤280 ms @ 50 ml/min (extended sampling
line)
Measured with a DRYLINE
TM
PRIME sampling line: ≤200 ms @ 50 ml/
min
Effect of interference gases on CO
2
measurement
Gas
Concentration (%)
Quantitative effect*
N
2
O
≤ 60
±1 mmHg
Hal
≤ 4
Sev
≤ 5
Iso
≤ 5
Enf
≤ 5
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...