A - 12
A.6.2.3
PR from Nellcor SpO
2
Module
A.6.2.4
PR from N
IBP
Module
A.6.3
NIBP
1
The Masimo pulse oximeter has been validated for low perfusion accuracy in bench top
testing against a Biotek Index 2 simulator and Masimo's simulator with signal strengths of
greater than 0.02% and a % transmission of greater than 5% for saturations ranging from 70
to 100%. This variation equals plus or minus one standard deviation. Plus or minus one
standard deviation encompasses 68% of the population.
Measurement range
20 to 300 bpm
Resolution
1 bpm
Response time
≤30 s (normal perfusion, no disturbance, PR value
sudden change within 25 – 250 bpm)
Accuracy
20 to 250 bpm: ±3 bpm
251 to 300 bpm: not specified
Refreshing rate
≤ 1 s
Measurement range
30 to 300 bpm
Resolution
1 bpm
Accuracy
±3 bpm or ±3%, whichever is greater
Standards
Meet standards of IEC 80601-2-30, ISO 81060-2
Technique
Oscillometry
Mode of operation
Manual, Auto, STAT, Sequence, and BP averaging
Auto mode repetition
intervals
1 min, 2 min, 2.5 min, 3 min, 5 min, 10 min, 15 min, 20 min, 30
min, 1 hour, 1.5 hours, 2 hours, 3 hours, 4 hours, 8 hours
STAT mode cycle time
5 min
Typical measurement
time
Adult: ≤ 15 s (Inflation algorithm, measured with CM1203/
CM1303/CM1503 cuff, PR within 60 to 200 bpm and systolic
pressure within 80 to 120 mmHg)
Pediatric: ≤ 15 s (Inflation algorithm, measured with CM1202/
CM1302/CM1502 cuff, PR within 60 to 200 bpm and systolic
pressure within 80 to 120 mmHg)
Max measurement time
Adult, pediatric:
180 s
Neonate: 90
s
Heart rate range
30 to 300bpm
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...