15 - 7
◆
On
: you need to confirm that whether the scoring result is saved or not after
the scoring is completed.
15.1.6.3
Setting the Manual Data Timeout
The manually input parameter data become invalid after a preset time. To set the
timeout period for the input data, follow this procedure:
1.
From the EWS screen select
Setup
.
2.
From the
Manual Data Timeout
area, select a desired parameter and set its
timeout period.
NOTE
•
If the data is expired and not updated, the monitor displays the
corresponding parameter score in outline font, and gives a timeout alarm.
15.1.6.4
Selecting the Default Scoring Tool
To set the default scoring tool for different patient catalog, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select .
2.
Select the
CAA
tab
→
select
EWS
tab.
3.
From the
Select Default Score
area, set
Default Adult Score
,
Default Ped Score
,
and
Default Neo Score
.
15.1.7
Viewing History Scores
To view the history scores, follow this procedure:
1.
Select the
Review
quick key
→
select
Scoring Review
tab.
2.
Select
PID
and from the popup list select the ID of which the scores are to be
viewed.
3.
Select Filter and choose a protocol if needed.
4.
Select a record and then select
Detail
.
15.1.7.1
Managing the Scoring Tools
To manage the scoring tool, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
System
column select
Maintenance
→
input the required password
→
select .
2.
Select the
CAA
tab
→
select
EWS
tab.
3.
Select
Manage Score
.
◆
Delete local scoring tools: from the
Local
page, delete the unnecessary
scoring tools.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...