13 - 3
3.
Switch on
RR
.
13.5.3
Setting RR Alarms
In
Continuous Monitoring
mode, you can set the RR alarm limits. Follow this procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2
menu.
2.
Select
Alarm
tab.
3.
Set the high and low limits and the priority in the
RR
row.
13.5.4
Setting the Apnea Alarm Delay
The monitor will alarm if the patient has stopped breathing for longer than the preset
Apnea time.
In
Continuous Monitoring
mode, you can set the Apnea delay time, follow this
procedure:
1.
Select the CO
2
numeric area or waveform area to enter the
CO2 Setup
menu.
2.
Select
Apnea Delay
and then select the appropriate setting.
13.5.5
Setting Timer Interval
A timer is provided when manually inputting RR value. A reminder is given at a set
interval to assist in counting of Resp number. To set the timer reminding interval for
manual RR, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Configuration
column select
Workflow
→
input the required password
→
select
.
2.
Select
on the right of the workflow to be set.
3.
Select
Parameters Setup
→
RR
tab.
4.
Set
Timer Reminder Interval
as needed.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...