11 - 14
3.
Select
Parameters Setup
→
the
NIBP
tab.
4.
Under
Average
, enable
Discard First Group of Readings
.
11.10 Orthostatic BP Measurement
Orthostatic blood pressure is important for the assessment of fall risks. On the monitor,
you can measure the lying and standing blood pressure of a patient, and check the
differences for clinical evaluation. Orthostatic BP measurement is only applicable for
adult patient.
A license is required for the orthostatic BP measurement function. To check the license,
see
3.9.9 Checking Software Licenses
.
11.10.1 Enabling Orthostatic BP Measurement
Orthostatic BP measurement is available only when the monitor is working in
Spot
Check
mode.
To enable Orthostatic BP measurement, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Configuration
column select
Workflow
→
input the required password
→
select
.
2.
Select
on the right of the workflow to be set. Make sure the workflow is
under
Spot Check
.
3.
Select
Parameters Setup
→
the
NIBP
tab.
4.
Under
Orthostatic BP Measurement
, enable
Orthostatic BP Measurement
.
11.10.2 Orthostatic BP Measurement Display
The following figure shows the orthostatic BP measurement screen.
1.
NIBP unit
2.
Lying blood pressure: displayed in the form of Sys/Dia (Mean)
3.
Standing blood pressure: displayed in the form of Sys/Dia (Mean). The lowest val-
ues measured are used in contrasting.
4.
Differences between the lying blood pressure and lowest standing blood pressure
1
3
2
5
8
4
7
6
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...