11 - 11
■
Patient having frequent arrhythmia
■
Weak patient pulse
NOTE
•
The monitor always performs NIBP measurement at deflation stage if Patient
Category is Neo.
11.8.8
Setting Measurement Times
To set the measurement times, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Configuration
column select
Workflow
→
input the required password
→
select
.
2.
Select
on the right of the workflow to be set. Make sure the workflow is
under
Spot Check
.
3.
Select
Parameters Setup
.
4.
Under
NIBP
tab, set
Measurements
to
Twice
.
11.8.9
Correcting the NIBP Measurements
The middle of the cuff should be at the level of right atrium. If the limb is not at the heart
level, you need to correct the measurement:
■
Add 0.75 mmHg (0.10 kPa) to the displayed value for each centimetre higher.
■
Deduct 0.75 mmHg (0.10 kPa) to the displayed value for each centimeter lower.
11.9
BP Averaging
In BP averaging mode, a group of BP measurement is performed on a patient
automatically. At the completion of measurement, a group of averaged BP readings is
displayed.
11.9.1
Enabling BP Averaging
BP averaging is available only when the
Department
is set to
Physician Office
and the
monitor is working in
Spot Check
mode.
To change the department, see
.
To enable BP averaging, follow this procedure:
1.
Select the
Main Menu
quick key
→
from the
Configuration
column select
Workflow
→
input the required password
→
select
.
2.
Select
on the right of the workflow to be set. Make sure the workflow is
under
Spot Check
.
3.
Select
Parameters Setup
→
the
NIBP
tab.
Summary of Contents for VS 8A
Page 1: ...VS 9 VS 9A VS 9C VS 8 VS 8A VS 8C Vital Signs Monitor Operator s Manual ...
Page 2: ......
Page 8: ...2 6 This page intentionally left blank ...
Page 26: ...1 6 This page intentionally left blank ...
Page 58: ...3 24 This page intentionally left blank ...
Page 66: ...5 4 This page intentionally left blank ...
Page 70: ...6 4 This page intentionally left blank ...
Page 98: ...8 16 This page intentionally left blank ...
Page 142: ...13 4 This page intentionally left blank ...
Page 162: ...15 16 This page intentionally left blank ...
Page 166: ...16 4 This page intentionally left blank ...
Page 202: ...19 8 This page intentionally left blank ...
Page 224: ...22 10 This page intentionally left blank ...
Page 250: ...B 8 This page intentionally left blank ...
Page 260: ...C 10 This page intentionally left blank ...
Page 271: ...D 11 This page intentionally left blank ...
Page 272: ...D 12 ...
Page 275: ...F 1 F Declaration of Conformity ...
Page 276: ...F 2 This page intentionally left blank ...
Page 277: ......
Page 278: ...P N 046 020097 00 4 0 ...