6 - 15
Test and Calibration
Testing the Optional Temperature Module
6.7
Testing the Optional Temperature Module
Required Tool: Thermostatic oil tank, HART 7102 recommended
1.
Set the temperature of the oil tank to 37
°
C and conduct the test after the temperature
stabilizes.
2.
Press
to display the
SYSTEM SETUP
dialog as shown in FIGURE 6-13.
FIGURE 6-13 SYSTEM SETUP
dialog
3.
Press
or
to highlight
TEMP SETUP
.
4.
Once
TEMP SETUP
is highlighted, press
to display the
TEMP SETUP
dialog as
FIGURE 6-14
Example
TEMP SETUP
dialog
5.
Press
or
to highlight the
TEMP TYPE
pull-down list.
6.
Once the pull-down list is highlighted, press
to view the selections.
7.
Press
or
to change the selection to
MONITOR
.
8.
Once MONITOR is highlighted, press
to set it.
9.
Press
or
to highlight the
TEMP UNIT
pull-down list.
10.
Once the pull-down list is highlighted, press
to view the selections.
NOTE:
For this test, the monitor and test tank temperature units are
in Celsius.