Electrical Safety Inspection E-1
Appendix E Electrical Safety
Inspection
The following electrical safety tests are recommended as part of a comprehensive preventive
maintenance program. They are a proven means of detecting abnormalities that, if undetected, could
prove dangerous to either the patient or the operator. Additional tests may be required according to local
regulations.
All tests can be performed using commercially-available safety analyzer test equipment. These
procedures assume the use of a 601PROXL International Safety Analyzer or equivalent safety analyzer.
Other popular testers which comply with IEC 60601-1 and are used in Europe, such as Fluke, Metron or
Gerb, may require modifications to the procedure. Follow the analyzer manufacturer's instructions.
An electrical safety inspection should be periodically performed every two years.
The safety
analyzer is also an excellent troubleshooting tool for detecting abnormalities in line voltage and
grounding, as well as total current loads.
E.1
Power Cord Plug
E.1.1 The Power Plug
Test Item
Acceptance Criteria
The power plug
The power plug pins
No broken or bent pins. No discolored pins.
The plug body
No physical damage to the plug body.
The strain relief
No physical damage to the strain relief. No plug
warmth when device is in use.
The power plug
No loose connections.
The power cord
No physical damage to the cord. No deterioration
to the cord.
For devices with detachable power cords, inspect
the connection with the device.
For devices with non-detachable power cords,
inspect the strain relief at the device.