Passport V Operating Instructions
10 - 7
CO
2
Monitoring (optional)
CO
2
Menu
10.3.2
Microstream
®
CO
2
Monitoring (Optional)
Microstream CO
2
modules provide
Expired CO
2
,
Inspired CO
2
and
Respiration Rate
monitoring utilizing a small lumen FilterLine
®
. Microstream capnography is accomplished via
a nasal cannula (non-intubated) or through an adapter set for use in a breathing circuit
(intubated). Microstream can be used on adult, pediatric and neonatal patients.
WARNING: When monitoring an anesthesized patient in an operating
To begin monitoring Microstream CO
2
:
1.
Open CO
2
input door and connect the proper FilterLine
®
to the Monitor. Connect the
opposite end to the patient.
NOTE:
Ensure all tubing connections are secure.
NOTE:
Consideration should be given to the disposal of packaging
waste.
CAUTION:
When connecting a sampling line to the monitor, screw the
sampling line connecter clockwise into the monitor CO
2
port
until it can no longer be turned to ensure that it is connected
securely to the monitor. This will assure that there is no leak
of gases during measurement at the connection point and
that measurement accuracy is not compromised. Loose or
damaged connections may compromise ventilation or cause
an inaccurate measurement of respiratory gases. Securely
connect all components and check connections for leaks
according to standard clinical procedures.
CAUTION:
If too much moisture enters the sampling line (i.e., from
ambient humidity or breathing of unusually humid air), the
message “Clearing FilterLine” will appear in the message
area. If the sampling line cannot be cleared, the message
“FilterLine Blockage” will appear in the message area.
Replace the sampling line once the “FilterLine Blockage”
message appears.
N
2
O
Compensation
0
(default) to 100%
Select the N
2
O level used for N2O
compensation.
Note
: The sum of compensation values input for
oxygen, nitrous oxide and desflurane must not
be greater than 100%.
Desflurane
Compensation
0
(default) to 24%
Select the Des level used for Des compensation.
Note
: The sum of compensation values input for
oxygen, nitrous oxide and desflurane must not
be greater than 100%.
CO
2
Setup Menu (DPM Sidestream Only)
MENU ITEM
SELECTIONS
COMMENTS
Summary of Contents for Passport V
Page 1: ...Operating Instructions Passport 0070 01 0704 02_PPV ops color indd 1 2 17 11 4 23 PM ...
Page 2: ...Operating Instructions Passport 0070 02 0704 02_PPV ops b_w indd 1 2 17 11 4 32 PM ...
Page 23: ...xx 0070 0 0704 02 Passport V Operating Instructions This page intentionally left blank 0 ...
Page 286: ...This page intentionally left blank ...