17 - 74
Operator’s Manual
17 Probes and Biopsy
4.
Immerse the needle-guided bracket fully in the sterilant and shake the bracket appropriately to
remove any bubbles on the surface. Use a syringe to draw an appropriate amount of sterilant
and inject the sterilant into the hole to remove the bubbles inside the hole if necessary.
For details about the immersion duration, see the operator’s manual provided by the
manufacturer.
5.
After sterilization, wash the needle-guided bracket thoroughly by using a large amount of
sterile distilled water (about 2 gallons) at room temperature for about 30 s to remove the
residual sterilant. Repeat the operation twice.
6.
Dry the needle-guided bracket by using a piece of sterile disposable lint-free soft cloth.
7.
Store the needle-guided bracket in a cool, clean and dry environment.
High-pressure steam sterilization
Perform the following procedure:
1.
Wear a pair of gloves to prevent infection.
2.
Clean thoroughly in accordance with the cleaning procedure before sterilization.
3.
Package the needle-guided bracket in accordance with the sterilization requirements of
surgical instruments.
4.
Place the packaged needle-guided bracket inside a high-temperature steam sterilizer and
perform sterilization.
The sterilization parameters are 121 °C and 30 minutes for a gravity displacement steam
sterilizer and are 132 °C and 4 minutes for a dynamic-air-removal steam sterilizer.
5.
Take out the sterilization package after sterilization and dry it in an oven at 60 °C for 20
minutes to 30 minutes.
Keep the sterilization package together with other sterilized surgical instruments in a sterile
item storage area.
17.2.7 Storage and Transportation
•
Do not use the carrying case for storing the needle-guided bracket. If the carrying case is used
for storage, it may become a source of infection.
•
Between examinations, keep the needle-guided bracket in a sterile environment.
•
If the needle-guided bracket is sent to your MINDRAY representative for repair, be sure to
disinfect or sterilize it and keep it in the carrying case to prevent infection.
•
Sterilize the carrying case as necessary.
•
Store or transport the needle-guided bracket under the following ambient conditions:
–
Ambient temperature: -20 °C to 55 °C
–
Relative humidity: 20% to 95% (no condensation)
–
Atmospheric pressure: 700hPa
-
1060hPa
17.2.8 Disposal
Be sure to sterilize the needle-guided bracket before disposing of it.
Contact your MINDRAY representative when disposing of this device.
Summary of Contents for Imagyn 7
Page 2: ......
Page 14: ...This page intentionally left blank...
Page 20: ...This page intentionally left blank...
Page 54: ...This page intentionally left blank...
Page 72: ...This page intentionally left blank...
Page 118: ...This page intentionally left blank...
Page 126: ...This page intentionally left blank...
Page 196: ...This page intentionally left blank...
Page 240: ...This page intentionally left blank...
Page 280: ...This page intentionally left blank...
Page 298: ...This page intentionally left blank...
Page 406: ...This page intentionally left blank...
Page 416: ...This page intentionally left blank...
Page 491: ......
Page 492: ...P N 046 019593 01 3 0...