7 3D/4D
Operator’s Manual
7 - 35
7.15.1 Basic Procedures for Smart Planes CNS
Perform the following procedure:
1.
Acquire necessary 3D data.
2.
Tap [S-Planes CNS]. There are four views detected by the system, which includes TCP, TTP,
MSP and TVP.
3.
Adjust the parameters to obtain the image.
4.
Tap [MSP]. There are A/B/C images appearing on the screen. The sagittal plane of the fetus
appears on window C.
Rotate A, B or C plane respectively, or rotate <4D> to adjust each plane.
NOTE:
To ensure the correctness of the result, please make sure the cerebral midline on the
anatomical plane is displayed clearly and ROI enclose the whole cranial region.
5.
Tap [OK] to accept the edit to the MSP. The system recalculates the TCP, TTP and TVP
according to MSP’s position. The position and the angle for TCP, TTP and TVP appear on
MSP plane.