Managing Patients
5.1
Discharging a Patient
Before monitoring a new patient, discharge the previous patient. After the patient is discharged, the technical
alarms is reset, and monitor settings return to their defaults. For more information, see
.
After a patient is discharged, the monitor automatically admits a new patient.
WARNING
•
Always discharge the previous patient before starting monitoring a new patient. Failure to do so can
lead to data being attributed to the wrong patient.
5.1.1
Manually Discharging a Patient
To manually discharge a patient, choose any of the following ways:
■
Swipe down the touchscreen with two fingers.
■
Select the
Discharge Patient
quick key.
■
Select the patient information area at the top left corner of the screen
→
Discharge Patient
.
■
Select the
Patient Management
quick key
→
Discharge Patient
.
■
Select the
Main Menu
quick key
→
from the
Patient Management
column select
Discharge Patient
.
Select the desired item from the popup box:
■
Print End Case Report
: prints the end case report when the patient is discharged.
■
Discharge
: clears the waveform data of the current patient. The monitor loads the default configuration
and goes to the standby mode. The current patient becomes a discharged patient.
■
Clear Patient Data
: discharges the current patient and clears the waveform data. The monitor still uses the
current configuration and does not go to the standby mode. The current patient becomes a discharged
patient.
5.2
Admitting a Patient
The monitor admits a new patient in the following situations:
■
After a patient is manually discharged, the monitor automatically admits a new patient.
■
After being switched off for the selected time period, the monitor automatically discharges the previous
patient and admits a new patient at startup.
■
If the monitor has not detected certain patient vital signs (ECG, SpO2, PR, RR, NIBP) for 30 minutes, you will
be prompted whether to start monitoring a new patient if any of the above vital signs are detected again.
Always inputs patient information as soon as the patient is admitted. For more information, see
for details.
Summary of Contents for ePM 10M
Page 1: ...ePM 10M ePM 10MA ePM 10MC ePM 12M ePM 12MA ePM 12MC Patient Monitor Operator s Manual ...
Page 2: ......
Page 58: ...4 8 This page intentionally left blank ...
Page 62: ...5 4 This page intentionally left blank ...
Page 118: ...11 4 This page intentionally left blank ...
Page 134: ...13 12 This page intentionally left blank ...
Page 144: ...15 8 This page intentionally left blank ...
Page 156: ...16 12 This page intentionally left blank ...
Page 174: ...18 12 This page intentionally left blank ...
Page 182: ...19 8 This page intentionally left blank ...
Page 192: ...20 10 This page intentionally left blank ...
Page 222: ...24 4 This page intentionally left blank ...
Page 228: ...25 6 This page intentionally left blank ...
Page 256: ...28 6 This page intentionally left blank ...
Page 264: ...29 8 This page intentionally left blank ...
Page 268: ...30 4 This page intentionally left blank ...
Page 280: ...31 12 This page intentionally left blank ...
Page 346: ...E 4 This page intentionally left blank ...
Page 350: ...F 4 This page intentionally left blank ...
Page 360: ...G 10 This page intentionally left blank ...
Page 361: ...H 1 H Declaration of Conformity ...
Page 362: ...H 2 This page intentionally left blank ...
Page 363: ......
Page 364: ...P N 046 012607 00 6 0 ...